Medicom MTD logo Medical equipment for electrodiagnosis, rehabilitation and research
 
PRODUCTS
Medicom-MTD. Electroencephalographic analyser
ENCEPHALAN 131-03 Electroencephalographic Analyser
  • A family of Electroencephalograph-Analysers: with 8, 16, 19 or 24 EEG-inputs (32 digital channels and 2marker channels) plus photo-, video-, audio-, and electric stimulation and patterned video stimulation outputs. Optionally, up to 6 polygraphical (ECG, SCG, SGR, EMG, EOG, PPG, CHD, RESP, TEMP, etc.) inputs and up to 6 timeshared rheoencephalographical (REG) inputs
     
  • Hospital or mobile system design with battery supply for emergency care and extremal situations service
     
  • High-quality data analysis and representation under Windows XP. Continuous recording and check-up of ultra-slow scalp potentials and electrode impedances
     
  • Convenient tools of visual EEG analysis including re-montage, split-screen review of EEG data, fast data scrolling, automatic artefact detection and rejection capabilities, 2D/3D mapping of EEG parameters in the Toposcope mode, etc
     
  • Comprehensive mathematical signal processing and analysis including spectral analysis (autocorrelation, cross correlation, coherence, etc.) and statistical analysis, with topographical mapping of all data
     
  • Analysis of functional asymmetry of EEG parameters with estimation of the degree of intra- and interhemispheric relationships
     
  • Automatical finding of epileptiform and transient events with marking-up of the transitory segments in long-term monitoring studies
     
  • A unique opportunity of concurrent recording and analysis of electroencephalograms (EEG), impedance plethysmograms (REG) and ultra-slow scalp activity (electrode impedances and slow potentials).
     
  • Automatical report generating with classification of EEG activity and manual editing of the report
     
  • 3D source localization of abnormal EEG sites
     
  • Late-latency EPs, somatosensory EPs, cognitive EP (CNV and P300), MMN and pattern stimulation studies
     
  • Long-term Video/EEG monitoring of a patient's state
     
  • Sleep EEG studies and objective psychophysiological testing intended for occupational care and psychophysiological research, based on an optional in-depth analysis of the polygraphical signals
     
  • Electronic Card Index (patient database) for running, editing and printout of patient and hospital documents, with convenient data exchange via a hospital information system
     
  • The Electroencephalograph-Analyser is intended for functional diagnosis, neurology and psychiatry departments, post-operative recovery rooms and neurological intensive care units in the neural diagnostic and rehabilitation centres and hospitals in a wide scope of clinical, educational and research applications
Complete set of "Encephalan 131-03" (depending on version/modification)
Encephalograph-131-03 Complete Set
  • Patient Box: a recording box with a corresponding set of inputs.
     
  • Desktop Holder or Mount Pillar for fixing a patient box/peripherals.
     
  • IB-2-USB Interface Unit (communication unit between a patient box and PC).
     
  • Software-controlled Photo and Audio Stimulators (2 stimulating channels each), Electrostimulator, Videostimulator and 3-button Remote Patient Terminal.
     
  • Set of EEG, ECG, EMG, EOG, SGR, PPG, and PG electrodes and accessories.
     
  • Application Software, Procedure Guide, and Help System under Windows XP
     
  •  Pentium IV 2 GHz or better, 17" Monitor, Printer (Notebook PC for mobile applications)
Specifications
Patient Box
  • Digital Signal Processing: 22-bit ADC, RS-133, ADSP-2181 microprocessor
     
  • Sensitivity: 0.1 - 200.0 uV/mm (in 21 steps) for EEG and EP; 0.02 - 5.0 mV/mm (in 8 steps) for Ultra- Slow Potentials; 0.02 - 1.0 kOhm/mm (in 21 steps) for Electrode Impedances
     
  • Time Constants: 10.0, 3.0, 1.0, 0.3, 0.1, 0.03, and 0.01 sec
     
  • High Frequency Filter (HFF): 0.016, 0.05, 0.16, 0.5, 1.6, 5.0, and 16.0 Hz
     
  • Low Frequency Filter (LFF): 5.0, 15.0, 30.0, and 70.0 Hz
     
  • AC Rejection Filter (50/60 Hz): 1:1000 or better
     
  • Automatic suppression of myographical artefacts: 20 dB (30 Hz).
     
  • Ultra-low Noise Level: 0.9 uV(p-p) max
     
  • Common Mode Rejection Ratio (CMRR): 125 dB or better
     
  • Patient Isolation: MEC601-1, Class I, Type BF
     
  • Calibration: Square (1 Hz) or Sine (5 Hz) signal of 5, 10, 20, 50, 100, 200, 500, 1000, 2000, 4000 uV
     
  • REG Channel Sensitivity:
          Total Impedance: 5.0 - 500.0mOhm/mm(7 steps),
          First Derivative: 0.1 - 10.0 (mOhm/sec)/mm (7 steps)
     
  • Polygraphic Channel Sensitivity: 0.1 - 200.0 uV/mm (in 13 steps).
“Encephalan-131-03” Software Versions

  • Basic Version. Routine EEG studies with computerized analysis and topographical mapping. Optionally, rheoencephalographical studies
     
  • Optimal Version. In-depth analysis of EEG/EP activity, analysis of epileptiform activity, differential diagnosis, long-term EEG monitoring, combined assessment of EEG and REG data using integral estimation of the brain activity and cerebral blood flow with REG-based topographical mapping of cerebral hemodynamic indices
     
  • Professional Version. Total analysis of cerebral activity (EEG, REG and ultra-slow scalp activity) with estimation of the degree of intra- and interhemispheric relationships
     
  • Elite Version. Advanced capabilities for carrying out EEG studies by using state-of-the-art techniques of electrophysiological data processing with total analysis of EEG, polygraphical, and ultra-slow cerebral activity. Automatical detection and suppression of ECG, EMG and EOG artefacts, mathematical processing and analysis of statistical data, data export in MatLab, EDF/UDF, and ASCII formats with a diverse spectrum of intelligent servicing tools
     
  • Elite-REG Version Same as "Elite" but with a full set of rheoencephalographic studies, concurrent EEG and REG visual analysis and integral assessment of electroencephalographic activity and cerebral hemodynamics
Certification and Patents